Testron 2400
A 2KW/2.4KW programmable
| |
multiple-output DC Power Supply System
| |
Download brochure in pdf format 1.17MB
|
Features
|
|
Fully GPIB & RS232 controlled
| |
|
True single-box solution
| |
|
19" by 51/4" (3U) mainframe, up to 8 modules
| |
|
Modules available in 5 power levels
| |
|
Individual module processor control
| |
|
Two power envelope levels
| |
|
Power envelope monitoring
| |
|
Expandable system (up to 16 modules)
| |
|
Modules built-in polarity and isolation relays
| |
|
Readback of voltage and current
| |
|
External synchronization
| |
|
Extensive D.U.T. protection features
| |
|
Workpoint window warning
| |
|
Multi-channel 10 store/recall locations
| |
|
99 steps auto sequencing
| |
|
Easy, software-based calibration
| |
|
Local panel and keyboard
| |
|
Low ripple and noise
| |
|
Power factor correction, wide range mains operation
| |
|
Fast response
| |
|
CE and UL approvals
|
|
|
|
Description
Easy Programming
----------------------------
By using a simple "English-like" set of commands, the Testron 2400 is easily programmed via a GPIB or RS-232C communication link, even for the novice user.
A well-structured extended command set covering the many new features of the Testron 2400 and compliance with the IEEE488.2 standard provide a familiar and easy to use instrument to the experienced user.
Programmable Functions
------------------------------------
Voltage and Current Setting.
O.V. and O.C. Protection.
Individual and Global Output ON/OFF.
Modules Built-in Output Relays.
Five current Limit Types.
Work-point Windows Warning.
Reprogramming Delay.
Local Panel Lockout.
Power-on Values Retain or Initialize.
Multi-channel Settings Store and Recall.
Output Synchronization.
Maskable Events and SRQ Generation.
Selectable Response to Controllers Communications Time-out.
Versatile Modules Shut-Down-on-Fault Setups.
Easy, Software-based Calibration.
Two-level, Six Frequencies Simulated Ripple.
Readback Functions
------------------------------
Load Voltage, Output Voltage and Load Current Readback.
All Programmed Parameters.
System Status.
Channel Status.
System and Channel Errors.
Synchronization to Commands Execution.
ATE Controller Offloading
------------------------------------
Work Point Window Warning.
Multi-channel Settings Store and Recall.
Ramp Function.
Operation Complete Indication.
Automatic Polarity Reversal.
Auto Sequence for Stand-alone Operation.
Internal Operations Synchronization.
External Synchronization.
Protection Features
----------------------------
Extensive Power-up self-test.
On-the-fly self-test.
User initiated self-test.
5 Types of Current Limitation: current limit, linear foldback, non-linear foldback, retry, single event shut-down.
Automatic (or programmable) OV and OC protection.
Sense disconnect protection.
Output disconnect relays.
Selctable output group shut-down.
Programmable shut-down on non-fault events.
Controller communications monitoring.
Over heating protection.
Power envelope protection.
Mains voltage out-of-range protection.
|
Specification
| |
Input
|
|
| |
Voltage
|
170-265 VAC
| |
Frequency Range
|
45-66 Hz
| |
Inrush Current
|
up to 100% of nominal
| |
Type of Connection
|
Power Cord
|
Output
|
|
| |
No. of Outputs
|
User Selctable (1 to 8)
| |
Voltage/Current Rating
|
See modules
| |
Power Rating
|
2KW/2.4KW cont.
| |
Temperature Derating
|
1% per Deg C from 30 to 55
|
Enviromental
|
|
| |
Operation Temperature
|
0 to 55 Deg C
| |
Storage Temperature
|
-20 to 70 Deg C
|
Indicators
|
|
| |
Power On
|
LED
| |
Remote
|
LED
| |
Summary Fault
|
LED
| |
Detailed Faults
|
using alfa-numeric display
| |
Current Limit
|
using alfa-numeric display
| |
TLK/LSN
|
LEDs
|
Metering
|
|
| |
Digital
|
using alfa-numeric display
| |
Full Display
|
data is user selectable
|
Signals
|
|
| |
Dry Contacts(relays)
|
Operating
|
Protection
|
|
| |
Input Fuse
|
Circuit breaker
| |
Over Power
|
2 levels
| |
Over Voltage
|
Automatic / programmable
| |
Over Current
|
Automatic / programmable
| |
Current Limit
|
5 types
| |
Short Circuit
|
Yes (not inc. internal relays)
| |
Thermal
|
shut-down
| |
Sense
|
2 levels
|
Communications
|
|
| |
GPIB
|
Control
| |
RS232
|
Control and Monitor
| |
Horizon net
|
Expansion
|
General
|
|
| |
Approvals And Standards
|
CE and UL (see catalog)
| |
Type of Case
|
Metal, 19" rack
| |
Weight
|
11.9Kg (chassis)
| |
Size LxWxH
|
570x482.6x132.6 mm
|
|
|
| |